Subspace Learning with Interval-Censored Likelihoods for Dequantizing Percept PC LFP Snapshots
Implanted neurostimulators that sense local field potentials now enable chronic electrophysiology based biomarker tracking in patients at home. The Medtronic Percept PC, the only commercially available sensing-enabled deep brain stimulation (DBS) device, stores spectral amplitudes as 16-bit integers at approximately 0.1 $μ$V per bit (quantum $q \approx 0.11$ $μ$Vp). At frequencies where the true amplitude spans only a few quantization levels, consecutive bins round to the same stored value. Standard spectral parameterization (FOOOF, fitting oscillations and one over f), which separates periodi